Ann Thorac Surg 2002;74:2003-2009
© 2002 The Society of Thoracic Surgeons
a Department of Medicine, Division of Cardiology, University of Michigan, Ann Arbor, Michigan, USA
b Section of Cardiothoracic Surgery, Department of Surgery, William Beaumont Hospital, Royal Oak, Michigan, USA
c Section of Cardiothoracic Surgery, Department of Surgery, London Health Sciences Center, London, Ontario, Canada
d Section of Cardiothoracic Surgery, Department of Surgery, St. Thomas Hospital, Nashville, Tennessee, USA
e Section of Cardiothoracic Surgery, Department of Surgery, Herzzentrum Universitat Leipzig, Leipzig, Germany
* Address reprint requests to Dr Bach, University of Michigan, L3119 Womens0273, Ann Arbor, MI 48109, USA.
e-mail: dbach{at}umich.edu
Presented at the Thirty-eighth Annual Meeting of The Society of Thoracic Surgeons, Fort Lauderdale, FL, Jan 2830, 2002.
| Abstract |
|---|
|
|
|---|
METHODS: Between July 1998 and July 2001, 361 patients at 17 centers in North America and Europe underwent implantation of a Regent mechanical aortic valve prosthesis. Clinical status was prospectively recorded, and echocardiography with Doppler was performed at discharge and at 2 months, 6 months, 1 year, and 2 years after operation.
RESULTS: Follow-up to date is 300 patient-years (average, 0.8 ± 0.7 years per patient; range, 0.0 to 2.7 years). There were low rates of clinical adverse events. Mean gradient at 6 months was 9.7 ± 5.3 mm Hg, 7.6 ± 5.2 mm Hg, 6.3 ± 3.7 mm Hg, 5.8 ± 3.4 mm Hg, and 4.0 ± 2.6 mm Hg, respectively, for 19-mm, 21-mm, 23-mm, 25-mm, and 27-mm valves; effective orifice area was 1.6 ± 0.4 cm2, 2.0 ± 0.7 cm2, 2.2 ± 0.9 cm2, 2.5 ± 0.9 cm2, and 3.6 ± 1.3 cm2, respectively. Indexed effective orifice area was equal to or greater than 1.0 cm2/m2 for all valve sizes. Left ventricular mass index decreased significantly between early postoperative (165.9 ± 57.1 g/m2) and 6-month follow-up (137.9 ± 41.0 g/m2;
= -28.0 ± 49.1 g/m2; p < 0.0001).
CONCLUSIONS: The St. Jude Medical Regent aortic valve has excellent hemodynamics and early clinical results, with rapid and significant left ventricular mass regression. Long-term clinical assessment is ongoing.
| Introduction |
|---|
|
|
|---|
| Patients and methods |
|---|
|
|
|---|
Preoperative valve dysfunction was aortic stenosis in 182 (50.4%) patients, aortic regurgitation in 68 (18.8%), and mixed stenosis and regurgitation in 111 (30.7%). The cause of disease (more than one cause could be included for each patient) was degenerative in 212 (58.7%) patients, congenital in 124 (34.3%), rheumatic in 48 (13.3%), infective in 8 (2.2%), and other causes in 41 (11.4%). Surgical inspection of valve disease (patients could be counted more than once) revealed leaflet calcification in 260 (72.0%), annular calcification in 207 (57.3%), leaflet thickening in 102 (28.3%), commissural fusion in 84 (23.3%), annular dilation in 27 (7.5%), leaflet perforation in 12 (3.3%), and other diseases in 34 (9.4%). Aortic valve repair or replacement had been previously performed in 12 (3.3%) patients; prior coronary artery bypass grafting had been performed in 16 (4.4%). The distribution of aortic valve sizes implanted is shown in Figure 1. Concomitant surgical procedures were performed in 138 (38.2%) of 361 patients, and included coronary artery bypass grafting in 109 (30.2%) patients, mitral valve repair in 6 (1.7%), and aortic root enlargement in 4 (1.1%). Postoperative anticoagulation schemes were determined by individual physicians and study sites. The preponderance of sites used a target international normalized ratio range of 2.0 to 3.0 or 3.5.
|
|
Echocardiography
By protocol, transthoracic echocardiography with Doppler was performed at specified times, including before hospital discharge and at 2 months, 6 months, and 1 year postoperatively, and yearly thereafter. Imaging included two-dimensional and M-mode echocardiography, pulsed-wave and continuous-wave spectral Doppler, and color flow Doppler imaging. The aortic valve prosthesis was interrogated for regurgitation using parasternal and apical windows. Echocardiographic images were recorded on standard or super VHS videotape for subsequent analysis.
All echocardiograms were analyzed at a central, core echocardiography laboratory by observers highly trained in qualitative and quantitative echocardiography and Doppler. Observers had no knowledge of clinical information at the time of echocardiographic analysis. Echocardiographic analysis included two-dimensional echocardiographic measurement of LV dimensions and ejection fraction, measurement of mean and peak aortic gradients using the modified Bernoulli equation
[5], and calculation of aortic valve effective orifice area using the continuity equation [5]. Left ventricular mass was calculated using the modified American Society of Echocardiography cube method [6]:
![]() |
If present, aortic regurgitation was quantified as trivial, mild, moderate, or severe based on measurement of regurgitant jet diameter relative to the diameter of the LV outflow tract [7], assessment of aortic regurgitation deceleration slope on spectral continuous-wave Doppler, and assessment of diastolic flow reversal in the descending thoracic aorta. Aortic regurgitation when present was categorized by location as valvular, paravalvular, or indeterminate, based on jet location and characteristics in all available views.
Statistical analysis
All data are presented as mean ± standard deviation. Kaplan-Meier analyses for bleeding event, embolism, endocarditis, paravalvular leak, valve reoperation, all-cause mortality, and valve-related mortality were created using SAS system statistical software (SAS Institute, Inc, Cary, NC).
| Results |
|---|
|
|
|---|
|
|
|
Reoperation occurred in 3 patients within 2 weeks of operation and in 1 patient 14 months after operation. All reoperations were to address paravalvular leak; 2 (1 early and 1 late) were sequelae of infective endocarditis. Valves were replaced in 3 patients (2 early and 1 late), and the leak was repaired without valve replacement in the other.
Bleeding occurred in 32 patients (27 early and 5 late). Procedure-related bleeding occurred in 25 patients, and anticoagulation-related hemorrhage occurred in 10. (More than one bleeding event occurred in 8 patients, including both procedure-related and anticoagulation-related bleeding in 4.) Anticoagulation-related hemorrhage was gastrointestinal in 8, and intraocular, intracranial, and after ileostomy in 1 each. Anticoagulation was above target in four of eight cases of gastrointestinal bleeding, therapeutic in three, and unknown in one.
Hemodynamics
Echocardiograms were successfully performed in 318 (91.4%) of 348 eligible patients at discharge, 269 (84.2%) of 320 eligible patients at 2 months, 197 (77.9%) of 253 at 6 months, 120 (68.2%) of 176 at 1 year, and 39 (58.2%) of 67 at 2 years after operation. Values for mean pressure gradient, peak pressure gradient, and effective orifice area by valve size at 6 months are shown in Table 3.
|
= -1.4 ± 4.3 mm Hg; p = 0.001). Similarly, a statistically significant decrease in peak pressure gradient was observed between discharge (17.3 ± 7.9 mm Hg) and 6 months (14.1 ± 7.3 mm Hg;
= -3.2 ± 8.1 mm Hg; p = 0.0001). No further significant decrease in gradients was observed. Effective orifice area increased significantly between discharge (2.0 ± 0.8 cm2) and 6 months of follow-up (2.3 ± 0.9 cm2;
= +0.3 ± 1.0 cm2; p = 0.002). Cardiac output remained stable between early postoperative (5.9 ± 2.1 L/min), 6 months after operation (6.0 ± 2.1 L/min; p = 0.51), and 1 year after operation (6.3 ± 2.2 L/min; p = 0.34 versus early postoperative).
|
= -28.0 ± 49.1; p < 0.0001), with a large proportion of the decrease observed in the first 2 months. Between discharge and 6 months, LV mass decreased significantly for valve sizes 19 mm (135.3 ± 39.3 to 103.0 ± 16.0 g/m2; p = 0.03), 21 mm (148.2 ± 46.3 to 134.1 ± 41.0 g/m2; p = 0.006), 23 mm (172.4 ± 55.7 to 136.9 ± 31.8 g/m2; p < 0.0001), and 25 mm (176.1 ± 54.4 to 137.6 ± 47.5 g/m2; p = 0.003); LV mass regression was of similar magnitude but failed to reach statistical significance among patients with a 27-mm valve (202.1 ± 69.7 to 175.6 ± 61.7 g/m2; p = 0.26).
|
Aortic regurgitation
The presence and severity of aortic regurgitation on hospital discharge and subsequent follow-up intervals is shown in Figure 6. There was no to mild aortic regurgitation in 99.3%, 98.4%, 99.4%, 98.2%, and 96.9% of patients at discharge, 2 months, 6 months, 1 year, and 2 years, respectively. At 1 year after operation, more than mild aortic regurgitation was found in 2 patients: 1 with severe paravalvular regurgitation resulting from endocarditis, and 1 with moderate regurgitation of indeterminate origin. At 2 years, more than mild regurgitation occurred in a single patient, who had severe central valvular regurgitation. This patient remains in New York Heart Association functional class I, and no other tests have been performed to evaluate the valve.
|
| Comment |
|---|
|
|
|---|
Clinical outcomes
Ultimately, assessment of clinical outcomes relies on long-term assessment for years and decades. However, within the available follow-up of just more than 300 patient-years, there were excellent clinical outcomes among patients after implantation with the Regent mechanical aortic valve. Functional status assessed by New York Heart Association classification was excellent, and there were very low rates of adverse events. Postoperative bleeding events predominantly comprised early procedure-related events that were not directly related to the prosthesis. In all cases of adverse events, rates fell within objective performance criteria guidelines.
Hemodynamics in context
The present study reveals excellent hemodynamic variables associated with the Regent aortic valve prosthesis, with low mean and peak transvalvular gradients and large effective orifice area and indexed effective orifice area. Left ventricular mass regression was noted with smaller as well as with larger valve sizes, confirming that good transvalvular hemodynamics are associated with the entire range of available valve sizes. Although previously reported with stentless bioprostheses [8], this finding has not been described with mechanical prostheses.
Transvalvular gradients decreased and effective orifice area increased in the first 6 months after operation, which has not been previously reported with a mechanical prosthesis. Stable cardiac output and an increase in orifice area suggest that the decrease in gradients was not a reflection of altered flow across the valve. Incomplete valve opening early after operation is unlikely in light of normal cardiac output. The most likely explanation for improved hemodynamics in the months after valve replacement may be related to regression of LV outflow tract hypertrophy after aortic valve replacement, suggesting that early postoperative gradients were contributed to by subvalvular narrowing. Further research to test this hypothesis would be of interest.
The problem of prosthesis-patient mismatch is well documented and clinically evident among a subset of patients after aortic valve replacement [9, 10]. Prosthesis-patient mismatch results in less symptomatic improvement and suboptimal LV mass regression after aortic valve replacement [10]. Because prosthesis-patient mismatch of the aortic valve usually can be avoided if the indexed effective orifice area is more than 0.85 cm2/m2 [10], excellent hemodynamics in general, and the finding in the present report of indexed effective orifice area of at least 1.0 cm2/m2 for all Regent valve sizes, suggests that patient-prosthesis mismatch should be uncommon with the use of this prosthesis.
In the present report, there was a low prevalence of hemodynamically significant aortic regurgitation. Although trivial or mild aortic regurgitation was not uncommon, this is an anticipated finding with all bileaflet mechanical valves. There was a very low prevalence of significant periprosthetic regurgitation.
Study limitations
The present study reports clinical outcomes with follow-up of just more than 300 patient-years. The data are sufficient to conclude that there are excellent hemodynamics associated with the Regent valve. However, long-term follow-up is desirable to continue to assess clinical outcomes and objective performance criteria. No direct comparison is made between the Regent valve and other mechanical or biologic prostheses, although such comparisons may be the subject of future investigations.
Conclusion
The present report demonstrates excellent hemodynamic performance of the Regent bileaflet aortic valve prosthesis, with rapid and significant LV mass regression. Patient functional class and freedom from adverse events were very good. Long-term clinical assessment of the valve is ongoing.
| Appendix |
|---|
Canadian centers
United States centers
| Discussion |
|---|
|
|
|---|
DR EMERY: I would suspect so. There are sites that only stock the St. Jude Medical HP valve and do not stock the standard prosthesis. The Regent modification allows placement of a clearly larger valve size. It is very interesting because at times you think it might not fit, yet sizing is reliable.
DR JOACHIM LAAS (Bad Bevensen, Germany): I have two questions. We know from experimental and clinical studies that valve orientation, especially in small-size valves, is very important. I would like to know if you can give any advice on valve orientation?
The other question: you gain size by using a smaller sewing ring. However, is it really an advantage to implant a large size 27-mm valve and lose the safety of a pliable sewing ring? Furthermore, we know that in large size valves such as 27 mm, regurgitation can be as much as 25% and more in aortic and mitral subsitutes.
DR EMERY: First, I do not have the specifics of valve orientation from this study. Personally I orient the leaflets directed toward the left coronary ostium, with a leaflet guard at the left coronary, so they open in line with the artery. The Regent valve is rotatable as are the standard and HP St. Jude valves.
As far as the sizing of the valve, the sewing ring has less bulk. For implanting the standard St. Jude valve, which I have been using for a long time, I generally put in 12 or 13 mattress stitches. As the Regent sewing ring is smaller and thinner, I generally put in 2 or 3 more stitches and make the mattress sutures a bit narrower. I agree with you that there is possible risk. This has not turned out to be the case, however, neither in our hands nor in this multicenter study, that there is an increased incidence of paravalvular leak related to the sewing ring.
Regarding aortic regurgitation, the incidence is actually low.
(Slide) Here are the core laboratory echocardiographic results indicating none, mild, moderate, and trivial area. You can see over time that no significant aortic insufficiency occurs.
One patient had an explant for endocarditis because of a paravalvular leak that occurred late. This patient is the one who went from zero to mild to severe insufficiency with a central jet. She is in New York Heart Association class I. Neither the core laboratory nor the investigative site is perfectly clear what this really means. Originally the jet was called unknown on the echocardiogram. Then it was called a paravalvular leak. And at the last echocardiogram it was called central. The patient is fine, and we are just planning on following her along to find out how she does. We have not done anything.
So the regurgitation has really not been an issue with this prosthesis, and one can implant in an appropriate size that gives you excellent hemodynamics rather than trying to fit in an ultralarge size because you are concerned about gradients.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. Guenzinger, W. B. Eichinger, I. Hettich, S. Bleiziffer, D. Ruzicka, R. Bauernschmitt, and R. Lange A prospective randomized comparison of the Medtronic Advantage Supra and St Jude Medical Regent mechanical heart valves in the aortic position: is there an additional benefit of supra-annular valve positioning? J. Thorac. Cardiovasc. Surg., August 1, 2008; 136(2): 462 - 471. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. Emery, A. M. Emery, A. Knutsen, and G. V. Raikar Aortic Valve Replacement with a Mechanical Cardiac Valve Prosthesis Card. Surg. Adult, January 1, 2008; 3(2008): 841 - 856. [Full Text] |
||||
![]() |
P Pibarot and J G Dumesnil Prosthesis-patient mismatch: definition, clinical impact, and prevention Heart, August 1, 2006; 92(8): 1022 - 1029. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Dumesnil and P. Pibarot Prosthesis-patient mismatch and clinical outcomes: The evidence continues to accumulate J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 952 - 955. [Full Text] [PDF] |
||||
![]() |
J. S. Gammie, J. W. Brown, J. M. Brown, R. S. Poston, R. N. Pierson III, P. N. Odonkor, C. S. White, J. S. Gottdiener, and B. P. Griffith Aortic valve bypass for the high-risk patient with aortic stenosis. Ann. Thorac. Surg., May 1, 2006; 81(5): 1605 - 1610. [Abstract] [Full Text] [PDF] |
||||
![]() |
St Jude Medical Regent valve. J. Thorac. Cardiovasc. Surg., March 1, 2006; 131(3): 760 - 761. |
||||
![]() |
G. Tasca, Z. Mhagna, S. Perotti, P. B. Centurini, T. Sabatini, A. Amaducci, F. Brunelli, M. Cirillo, M. D. Tomba, E. Quiani, et al. Impact of Prosthesis-Patient Mismatch on Cardiac Events and Midterm Mortality After Aortic Valve Replacement in Patients With Pure Aortic Stenosis Circulation, January 31, 2006; 113(4): 570 - 576. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Otero, J. L. Pomar, J. M. Revuelta, and J. J. Rufilanchas Comparative Evaluation of Small-Size Sorin Slimline and St. Jude HP Heart Valve Prostheses Ann. Thorac. Surg., April 1, 2005; 79(4): 1284 - 1290. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. Emery, C. C. Krogh, K. V. Arom, A. M. Emery, K. Benyo-Albrecht, L. D. Joyce, and D. M. Nicoloff The St. Jude Medical Cardiac Valve Prosthesis: A 25-Year Experience With Single Valve Replacement Ann. Thorac. Surg., March 1, 2005; 79(3): 776 - 782. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Amarelli, A. Della Corte, G. Romano, G. Iasevoli, G. Dialetto, L. S. De Santo, M. De Feo, M. Torella, M. Scardone, and M. Cotrufo Left ventricular mass regression after aortic valve replacement with 17-mm St Jude Medical mechanical prostheses in isolated aortic stenosis J. Thorac. Cardiovasc. Surg., March 1, 2005; 129(3): 512 - 517. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Bach Choice of prosthetic heart valves:update for the next generation J. Am. Coll. Cardiol., November 19, 2003; 42(10): 1717 - 1719. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||