| Abstract |
|---|
|
|
|---|
Methods. Of 198 patients treated with a self-expanding polyester-covered stent-graft for various pathologic aortic conditions in our institution, we selected 70 consecutive patients with type B aortic dissection who were undergoing treatment. The stent-graft was introduced through the femoral artery in the angiography suite, under general anesthesia with systemic heparinization and induced hypotension.
Results. The procedure was performed in 70 patients; of these, 58 had descending aortic dissection and 12 had atypical dissections. The procedure was successful in 65 patients (92.9%), as documented by exclusion of the false lumen of the thoracic aorta. Eleven patients (18.9%) had persistent blood flow in the false lumen of the abdominal aorta due to distal reentries. Five patients (7.1%) underwent conversion to surgery. Insertion of additional stent-grafts was required in 34 patients (48.6%). At 29 months of follow-up, 91.4% of the patients were alive.
Conclusions. Stent-grafts are an important means of treating aortic dissections, which may replace conventional medical treatment of this condition for the majority of patients.
| Introduction |
|---|
|
|
|---|
The introduction of stent-grafts into clinical practice by Parodi [9] in 1995, as an alternative treatment for abdominal aortic aneurysms, led to new perspectives in the treatment of aneurysms and dissections at other aortic sites. Dake and colleagues [10] were the first to use stent-grafts in the thoracic aorta. Since then, many groups have reported success in treating various pathologic aortic conditions with different types of endoprosthesis, resulting in reduced rates of morbidity and mortality in selected group of patients [1113].
Clinical use of polyester-covered stent-grafts was pioneered at our institution for the treatment of type B dissections as a modification of the "elephant trunk" procedure [6]. The initial success in earlier phases of our experience encouraged us to develop new prostheses and catheters that could be deployed by means of a peripheral artery approach in the treatment of thoracic and abdominal aortic diseases. We report here the results and complications observed in a series of 70 consecutive, nonselected patients who presented with type B dissection, in whom a self-expanding endovascular prosthesis was used.
| Patients and methods |
|---|
|
|
|---|
15 days), whereas 12 patients had atypical dissections (6 patients had intramural hematomas and 6 had aortic wall ulcers). All patients presented with a complex clinical picture that included persistent pain, decreasing hematocrit levels, mediastinal hematoma, hemothorax, or hemoptysis. Comorbid conditions such as chronic obstructive pulmonary disease, aortic insufficiency due to annuloactasia, renal failure, cancer, heart failure, and coronary artery disease were commonly observed in this group. Diagnostic confirmation as well as the size and length of the required prosthesis were based on aortography, computed tomography (CT), and transesophageal echocardiography (TEE). Anatomical criteria required for stent-grafting were intimal tear in the descending aorta, proximal and distal landing zones up to 32 mm in diameter, and ilio-femoral segment able to accept a 20F device. The distance of the left subclavian artery from the entry tear was not a contraindication for the procedure. Seven patients underwent the procedure in critical condition. Two patients had had cardiopulmonary arrest upon admission to the hospital, resulting from hemoptysis due to bronchial aspiration. The other 5 patients had large hemothoraces secondary to leaks or rupture. Systemic hypertension was present on admission in 70% of the patients. Aortography revealed that in 18 patients (51.4%) the intimal tear was located less than 2 cm from the origin of the left subclavian artery. In 10 patients (28.9%) it was located 2 to 6 cm from the origin of the subclavian artery; in 7 patients (20%) the tear was present more distally in the descending portion of the thoracic aorta (Fig 1).
|
Endovascular prosthesis
The aortic stent-grafts were individually manufactured with self-expanding stainless steel stents covered with polyester (Braile Biomedica, SJ Rio Preto, São Paulo, Brazil). These devices are highly resistant to radial collapse and maintain their ability to return to their original diameter. The prostheses were deployed through a 20F catheter-based delivery system. Length and diameter varied according to the extension and size of the diseased segment to be treated. The stent-grafts were available in lengths of 7.5 cm and 9.0 cm; when fully expanded, their diameter ranged from 20 to 34 mm.
Operative procedure
The procedures were performed in the angiography suite under general anesthesia with endotracheal intubation and monitoring of vital signs. Transesophageal echocardiography (which is especially useful in cases of penetrating ulcers and intramural hematomas) was used as an important adjunct whenever possible. Aortography was performed in two views (ie, postero-anterior and left anterior oblique) using marked catheters to confirm preoperative CT-based measurements. After aortography, the least tortuous ileo-femoral artery was used to introduce the delivery catheter with the stent-graft. Surgical cut-down was used to expose the appropriate femoral artery. A dose of 5,000 U of heparin was used for anticoagulation. Under radiographic visualization, the catheter with the stent-graft was inserted over an extra-stiff guidewire through the femoral artery and slowly advanced toward the chosen deployment site in the thoracic aorta. The deployment site was marked by a radiopaque ruler placed under the patients back before the start of the procedure. Immediately before stent-graft deployment, sodium nitroprusside was administered intravenously to lower the mean aortic pressure to between 50 and 60 mm Hg, to decrease the risk of stent-graft migration during deployment. To confirm exclusion of the dissection, aortography was carried out a few minutes after the stent-graft was deployed and after blood pressure had returned to control values. After the procedure, patients were monitored in the postoperative care unit and kept under observation for 24 hours. The patients were discharged home after a postprocedural CT scan had been performed.
| Results |
|---|
|
|
|---|
The procedure was technically successful in 65 patients (92.9%), as documented by exclusion of the false lumen or ulcer/hematoma in the thoracic aorta (Fig 2). There were 5 (7.1%) conversions to surgery, all due to leakage at the proximal end of the stent-graft. Two patients required emergent conversion due to expansion of the aneurysm. Three patients underwent elective operations due to large flow of blood in the false lumen of the aneurysm. In 11 patients (18.9%) with classic type B dissection, residual blood flow in the false lumen persisted in the abdominal aorta due to a more distal tear, which could not be treated percutaneously because of the proximity of visceral vessels.
|
For adequate repair, insertion of additional stent-grafts was required in 34 patients (48.6%). Additional stent-grafts were needed in some patients either because of multiple tears or because the first stent-graft was malpositioned in relation to the origin of the dissection or the intimal tear. To that end, 17 patients received two stent-grafts, 10 received three, 4 received four, 2 received five, and 1 received six (Fig 3). Indications for additional stent-graft implantation were multiple entry tears or stent-graft malposition.
|
One case of brachial artery thrombosis (1.4%) was observed at an access site. This patient required endarterectomy repair. It was believed that this complication was due to the Seldinger technique used for insertion of the device. Another patient required endarterectomy repair of the femoral artery after the procedure. Another patient experienced deep vein thrombosis. Other complications were mild and transient renal failure (15 patients, 21.4%), surgical wound infection (2 patients, 2.9%), and peritoneal dialysis catheter infection (1 patient, 1.4%). Fever was common (15 patients, 21.4%); in most patients it appeared not to be related to an infective process and was considered to be due to an inflammatory reaction to the prosthesis. These patients were treated with antiinflammatory agents. Four patients (5.7%) died after hospital discharge. Two patients had sudden unexplained death. A third patient died in a car accident. The fourth patient died from hepatic failure. At mean follow-up of 29 months, 91.4% of the patients are alive and well.
| Comment |
|---|
|
|
|---|
In our series of patients, there were low mortality and surgical conversion rates. The two sudden deaths that were observed were likely related to aortic disease, demonstrating that these patients require very close follow-up. The presence of intimal tears in the abdominal aorta, maintaining blood flow between the lumens, was frequently not diagnosed until occlusion of the thoracic tears. Accurate diagnosis of abdominal intimal tears can only be made after proximal occlusion of the true and false lumens, thereby decreasing the pressure in the false lumen distally. During TEE it was evident that immediate thrombosis of the false lumen of the thoracic aorta had occurred, as also documented by predischarge CT.
Absence of paraplegia in this series of patients, despite the facts that some of them had multiple stent-grafts and some had more than two thirds of their descending thoracic aorta covered by stent-grafts, is very encouraging. The low incidence of paraplegia after stent-graft treatment of aortic disease has been reported by others [14].
Cerebral vascular accidents are potential complications and can be associated with stent-graft insertion, as the aorta is usually severely diseased. Minimizing manipulation of the catheters during the procedure may decrease the incidence of these complications.
Other complications occurring in our series of patients were minor. Fever, not related to infection, was frequent and responded to therapy with nonsteroidal anti-inflammatory drugs. It was postulated that fever was due to an immune-mediated process in patients presenting with low-grade fever and mild leukocytosis. The use of antibiotics must be judiciously used in this group of patients.
In conclusion, our successful initial experience in the treatment of complicated and uncomplicated type B dissection of the aorta with stent-grafts provides a less invasive and lower-risk approach than does conventional open surgery. The latter is associated with significant morbidity and mortality rates. We have demonstrated that it is possible to insert stent-grafts in the angiography suite with a low complication rate. However, experience with larger series of patients and longer follow-up periods, as well as refinements in stent-grafts and their delivery systems, are required before this new modality of therapy can be recommended for widespread use treating thoracic aortic dissections.
| Acknowledgments |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
Related Article
This article has been cited by other articles:
![]() |
J. D. Parker and J. Golledge Outcome of endovascular treatment of acute type B aortic dissection. Ann. Thorac. Surg., November 1, 2008; 86(5): 1707 - 1712. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Dunning, J. E. Martin, H. Shennib, and D. C. Cheng Is it safe to cover the left subclavian artery when placing an endovascular stent in the descending thoracic aorta? Interactive CardioVascular and Thoracic Surgery, August 1, 2008; 7(4): 690 - 697. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. Palma, J. S. Guilhen, D. F. Gaia, and E. Buffolo Early complication after hybrid thoracic aortic aneurysm repair Interactive CardioVascular and Thoracic Surgery, June 1, 2008; 7(3): 441 - 443. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. L. Kahn and M. D. Dake Stent Graft Management of Stable, Uncomplicated Type B Aortic Dissection Perspectives in Vascular Surgery and Endovascular Therapy, June 1, 2007; 19(2): 162 - 169. [Abstract] [PDF] |
||||
![]() |
M. Schoder, M. Czerny, M. Cejna, T. Rand, A. Stadler, G. H. Sodeck, R. Gottardi, C. Loewe, and J. Lammer Endovascular Repair of Acute Type B Aortic Dissection: Long-Term Follow-Up of True and False Lumen Diameter Changes Ann. Thorac. Surg., March 1, 2007; 83(3): 1059 - 1066. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Bockler, H. Schumacher, M. Ganten, H. von Tengg-Kobligk, M. Schwarzbach, C. Fink, H.-U. Kauczor, H. Bardenheuer, and J.-R. Allenberg Complications after endovascular repair of acute symptomatic and chronic expanding Stanford type B aortic dissections. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 361 - 368.e4. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Eggebrecht, C. A. Nienaber, M. Neuhauser, D. Baumgart, S. Kische, A. Schmermund, U. Herold, T. C. Rehders, H. G. Jakob, and R. Erbel Endovascular stent-graft placement in aortic dissection: a meta-analysis Eur. Heart J., February 2, 2006; 27(4): 489 - 498. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Dong Xu, Z. Zhong Li, F. J. Huang, J. F. Yang, X. Y. Wang, Z. G. Zhang, J. Hui Du, and Y. Q. Sun Treating Aortic Dissection and Penetrating Aortic Ulcer with Stent Graft: Thirty Cases Ann. Thorac. Surg., September 1, 2005; 80(3): 864 - 868. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Karck, A. Chavan, N. Khaladj, H. Friedrich, C. Hagl, and A. Haverich The frozen elephant trunk technique for the treatment of extensive thoracic aortic aneurysms: operative results and follow-up Eur. J. Cardiothorac. Surg., August 1, 2005; 28(2): 286 - 290. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Kusagawa, T. Shimono, M. Ishida, T. Suzuki, F. Yasuda, U. Yuasa, K. Onoda, I. Yada, T. Hirano, K. Takeda, et al. Changes in False Lumen After Transluminal Stent-Graft Placement in Aortic Dissections: Six Years' Experience Circulation, June 7, 2005; 111(22): 2951 - 2957. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Dialetto, F. E. Covino, G. Scognamiglio, S. Manduca, A. D. Corte, B. Giannolo, M. Scardone, and M. Cotrufo Treatment of type B aortic dissection: endoluminal repair or conventional medical therapy? Eur. J. Cardiothorac. Surg., May 1, 2005; 27(5): 826 - 830. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Nienaber, V. Skriabina, W. Schareck, and H. Ince To stent or not to stent aortic dissection: good news for a chosen few, but who? Eur. Heart J., March 1, 2005; 26(5): 431 - 432. [Full Text] [PDF] |
||||
![]() |
E. Therasse, G. Soulez, M.-F. Giroux, P. Perreault, L. Bouchard, J.-F. Blair, N. Beaudoin, A. Benko, and V. L. Oliva Stent-Graft Placement for the Treatment of Thoracic Aortic Diseases RadioGraphics, January 1, 2005; 25(1): 157 - 173. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Rampoldi, S. Trimarchi, P. Righini, V. Tolva, and L. Inglese Open aortic surgical repair for left hemi-arch stent-graft failure Ann. Thorac. Surg., September 1, 2004; 78(3): 1075 - 1078. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. A. Fabri, P. C. Santos, C. R. Cunha, L. Severino, V. J. Pereira, R. V. Botelho, J. H. P. Fonseca, and E. Buffolo Combined repair of extensive aneurysm of the ascending aorta, arch, and descending aorta Ann. Thorac. Surg., July 1, 2004; 78(1): 319 - 321. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Nienaber and K. A. Eagle Aortic Dissection: New Frontiers in Diagnosis and Management: Part II: Therapeutic Management and Follow-Up Circulation, August 12, 2003; 108(6): 772 - 778. [Full Text] [PDF] |
||||
![]() |
E. Buffolo, J. H. P. da Fonseca, J. A. M. de Souza, and C. M. R. Alves Revolutionary treatment of aneurysms and dissections of descending aorta: the endovascular approach Ann. Thorac. Surg., November 1, 2002; 74(5): S1815 - 1817. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Mitchell Stent grafts for the thoracic aorta: a new paradigm? Ann. Thorac. Surg., November 1, 2002; 74(5): S1818 - 1820. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |