Ann Thorac Surg 2000;69:1782-1786
© 2000 The Society of Thoracic Surgeons
a Division of Thoracic and Cardiovascular Surgery, Department of Surgery, The University of Texas Southwestern Medical Center at Dallas, Dallas, Texas, USA
b Department of Anesthesiology and Pain Management, The University of Texas Southwestern Medical Center at Dallas, Dallas, Texas, USA
c Division of Cardiology, Department of Internal Medicine, The University of Texas Southwestern Medical Center at Dallas, Dallas, Texas, USA
d Dallas Veterans Affairs Medical Center, Dallas, Texas, USA
Address reprint requests to Dr Jessen, Division of Thoracic and Cardiovascular Surgery, Department of Surgery, The University of Texas Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX 75235-8879
e-mail: michael.jessen{at}email.swmed.edu
| Abstract |
|---|
|
|
|---|
Methods. Thirty-four patients undergoing coronary artery bypass grafting were prepared by implanting a PA catheter, an EDM, and a transit-time ultrasonic flow probe around the ascending aorta. In 20 patients, left ventricular end-diastolic short-axis area (EDA) was measured by transesophageal echocardiography. At five time points, cardiac output was measured from the flow probe, the EDM, and the PA catheter (by thermodilution), and left ventricular filling was assessed from the PA catheter (as PA diastolic pressure), the EDM (corrected flow time), and the EDA. For cardiac output, concordance correlations relating EDM to flow probe and PA catheter to flow probe were calculated, transformed (Fishers z transformation), and compared by Students t test. For left ventricular filling, regression coefficients were created between corrected flow time and EDA and between PA diastolic pressure and EDA. Spearman correlations were compared by Wilcoxon rank sum test.
Results. The EDM and the PA catheter exhibited similar relationships to the flow probe (concordance correlations, 0.55 ± 0.35 [mean ± standard deviation] and 0.49 ± 0.34, respectively; p = 0.088). The correlation between corrected flow time and EDA was better than the correlation between PA diastolic pressure and EDA (concordance correlations, 0.49 ± 0.55 versus 0.10 ± 0.43, respectively; p < 0.01).
Conclusions. These data suggest that the EDM may offer a less invasive technique for evaluating cardiac output and a more accurate estimate for preload compared with the PA catheter.
| Introduction |
|---|
|
|
|---|
The most common tool in use is the pulmonary artery (PA) catheter [2]. This device provides a calculated CO determination from thermodilution, an indirect assessment of cardiac filling from PA diastolic or wedge pressures, and can supply a measurement of mixed venous oxygen saturation. These data can be important in the management of some patients in the early postoperative period, but PA catheters carry risks of major complications [3] and add substantial cost. The value of PA catheters has not been well defined in cardiac surgical patients. However, use of these devices may be associated with an increased mortality risk in patients with acute myocardial infarction [4] and in critically ill patients in the intensive care unit [5]. In cardiac surgical patients, the accuracy of PA catheters has also been questioned, as measurement of CO by thermodilution can be subject to error induced by rapid temperature shifts after separation from cardiopulmonary bypass (CPB) [6, 7].
An alternative method for monitoring cardiac performance uses Doppler signals acquired by an esophageal Doppler monitor (EDM). With this technique, a pulsed, competitive-frequency continuous wave Doppler signal is emitted from a probe placed in the distal esophagus and directed at the descending thoracic aorta. The reflected ultrasound frequencies are recorded from a transducer on the probe and are processed to display a continuously monitored blood velocity signal. These velocity data can then be converted to flow by applying an equation that includes the estimated cross-sectional diameter of the descending aorta, which is derived from nomograms [8]. The Doppler flow velocity waveform can also be analyzed to identify the flow time or duration of left ventricular ejection. This can be corrected for heart rate (cycle time) to yield an index of cardiac filling [9]. The device is simple to insert and, in initial studies, appears to yield CO measurements that correlate well with thermodilution values [10]. However, the technique is dependent on estimation of aortic diameter, its use can be affected by peripheral vascular disease, and it has seen limited application in patients undergoing cardiac operations [11].
The cardiac surgical patient provides a unique opportunity to compare these two monitoring devices. The surgical incision allows access to the ascending aorta for placement of a flow probe to directly measure left ventricular flow. Flow data can be compared with CO values simultaneously acquired from the thermodilution PA catheter and the EDM. In addition, transesophageal echocardiographic images can be obtained at the same setting to quantify left ventricular dimensions. These data can then be correlated with other indices of cardiac filling: PA diastolic pressure (PAD) measurements and Doppler-derived corrected flow time (FTc).
This study was undertaken to compare CO measurements and cardiac filling variables immediately before and after CPB in patients undergoing coronary artery bypass grafting.
| Material and methods |
|---|
|
|
|---|
|
An EDM (Deltex Medical, Irving, TX) was inserted through the oropharynx into the distal esophagus approximately 35 to 40 cm from the incisors. The probe was manipulated until an optimal signal was obtained. In 20 of the patients, a 5-MHz esophageal echocardiographic probe (Sonos 2500; Hewlett-Packard, Andover, MA) was also inserted into the midesophagus, and the heart was imaged.
A median sternotomy was performed, and saphenous vein and the left internal thoracic artery were harvested. The pericardium was incised and fashioned into a cradle. Each patient was administered heparin sodium in a dose of 300 U/kg intravenously, and an aortic cannula was placed in the distal ascending aorta and connected to the CPB circuit. A sterile aortic flow probe (Transonic Systems, Inc, Ithaca, NY) matched to the size of the ascending aorta (28 to 36 mm) was selected and placed around the ascending aorta proximal to the cannula. The probe was connected to a Transonic Systems flow computer and calibrated. The heart, aorta, and flow probe were immersed in warm, sterile saline solution. All pressure transducers were zeroed, and internal calibration of the mixed venous oxygen saturation monitor was completed.
The first set of data was then acquired. Heart rate, systemic and PA blood pressures, central venous pressure, and mixed venous oxygen saturation were recorded from their respective monitors. The EDM signal was optimized and the tracing frozen, and values for CO (COEDM) and FTc were recorded. The flow value from the aortic flow probe (COFP) was also recorded. A short-axis echocardiographic image of the left ventricle was obtained at the midpapillary level and recorded. Three separate thermodilution measurements of CO were obtained over 3 minutes by injection of 10 mL of room temperature saline solution into the proximal port of the PA catheter. If an unstable baseline message appeared, the resulting thermodilution value was discarded, and the process was repeated until a satisfactory measurement was obtained. After the three thermodilution measurements were made, a repeat measurement of COFP was recorded, and repeat values of COEDM and FTc were obtained.
Thermodilution injections and CO determinations were made by a senior anesthesiologist. The EDM data were acquired by a representative from the manufacturer (Deltex Medical). Echocardiographic images were acquired by a separate anesthesiologist or cardiologist familiar with the technique. All data were recorded by an individual investigator (MVC), and results of any one technique were blinded to investigators acquiring data by another method. Flow probe, EDM, and thermodilution measurements were averaged at each time point.
After initial data acquisition, volume from the CPB circuit (primed with 1,500 mL of lactated Ringers solution, 200 mL of 20% mannitol, 100 mL of 25% albumin, and 10,000 units of heparin) was infused through the aortic cannula until an increase in COFP of at least 1 L/min was observed. A repeat set of all variables was then measured and recorded. After this, additional volume was infused to achieve a further increment in flow, and all variables were measured and recorded again.
The aortic flow probe was removed, cannulation was completed, and the patient was placed on CPB and cooled to 28°C. Coronary artery bypass grafting was performed, and the patient was rewarmed during the construction of the proximal anastomoses. The lungs were reinflated, and the patient was separated from CPB. The aortic flow probe was repositioned, and signals from all devices were optimized.
Another set of measurements was recorded by the identical protocols 5 minutes after separation from CPB. Volume was then infused as judged appropriate by the surgical team who were observing the heart directly and monitoring systemic blood pressure. Fifteen minutes after separation from bypass, a final set of measurements was recorded. Thus, data were measured at a total of five time points for each patient. No protamine sulfate was administered until after the final time point. The aortic flow probe was removed and the operation completed.
Data analysis
For each patient at each time point, averages of CO measured from the two aortic flow probe recordings (COFP), the two EDM recordings (COEDM), and the three thermodilution recordings (COTD) were used. Each of the 34 subjects had five observations of CO (one from each time point) by the three different techniques. The direct measurement of aortic flow from the transit-time ultrasonic flow probe, COFP, was considered the gold standard. For each patient, a concordance correlation was computed for the relationship between COEDM and COFP and between COTD and COFP. Each concordance correlation was transformed to statistical normality by the Fisher z transformation. The 34 pairs of transformed concordance correlations were compared by a paired t test, and a p value of less than 0.05 was considered significant.
The comparison of indices of cardiac filling required a slightly modified analysis. The echocardiographic images were analyzed by tracing the left ventricular endocardium at end-diastole at the midpapillary level in the short-axis view to obtain an end-diastolic area (EDA). This measurement was considered the gold standard for ventricular filling [12]. Variables from the EDM (FTc) and the PA catheter (PAD) use indirect methods (time and pressure, respectively) to imply volume conditions. Therefore, unlike CO comparisons, correlations between these variables with different units cannot be analyzed with the assumption that values should be equal to the gold standard. Instead, filling measurements should rise or fall in proportion to changes observed in the EDA.
For each of the 20 patients with transesophageal echocardiographic data, Spearman correlations were computed for FTc versus EDA and for PAD versus EDA. These correlations examine how well the data fit the linear regression relating the two variables over the five time points as opposed to how closely they fit the equality line (y = x). The Spearman correlations of these regressions were compared by Wilcoxon rank sum test, and p values of less than 0.05 were considered significant.
| Results |
|---|
|
|
|---|
The relationship between COEDM and COFP for all patients at all time points is shown in Figure 1. In general, the correlation between COs measured by these two techniques was good. The equation of the linear regression of this relationship was calculated as: COEDM = 1.027 x COFP + 0.571. The relationship between COTD and COFP was also linear over the range studied. This relationship is depicted in Figure 2 and is described by the equation: COTD = 0.874 x COFP + 1.625.
|
|
For comparison of the indices of left ventricular filling, regressions were also created between FTc from the EDM and EDA from transesophageal echocardiographic information. Spearman correlation statistics from these regressions were compared with those calculated from the regression of PAD (from the PA catheter) and EDA. Regression coefficients (mean ± standard deviation) for FTc versus EDA and PAD versus EDA were 0.49 ± 0.55 and 0.10 ± 0.43, respectively. The regression coefficients for FTc were significantly higher than those for PAD by Wilcoxon rank sum test (p < 0.01).
| Comment |
|---|
|
|
|---|
Measurement of CO can be accomplished by a variety of techniques, each with unique advantages and disadvantages. Direct measurement of left ventricular output by an aortic flow probe is done by comparisons of antegrade and retrograde ultrasound transit-time and may offer the most accurate determination [15, 16]. However, this method requires invasive instrumentation that is not easily adapted to the postoperative arena, does not account for the portion of CO that supplies the coronary circulation, and can be inaccurate if the probe size does not match the ascending aorta precisely. Thermodilution CO determinations mandate intravascular and intracardiac line placement and can be affected by valvular abnormalities (especially tricuspid insufficiency [17]) and inaccuracies in injectate volume or temperature [18]. Thermal variations in patients recently rewarmed after CPB can also cause error [6, 7]. Thermodilution CO can be obtained continuously by some catheters currently available, but the initial measurements are not displayed until several minutes after separation from CPB, and therefore, standard injection methods were used in this study. The EDM is less invasive and simpler to insert, although esophageal instrumentation can entail some risks. The technique depends on the ability of the user to accurately focus the signal on the descending aorta and makes assumptions on aortic diameter based on algorithms from patient size.
In this study, EDM measurements of CO were at least as accurate as thermodilution determinations when compared with aortic flow measured directly. This suggests that the less invasive EDM device can play a useful role in monitoring the postoperative cardiac surgical patient. Thermodilution measurements were slightly, but not significantly, less accurate. The thermodilution CO measurements may have been adversely affected by the thermal changes induced by systemic rewarming after CPB. However, repeating the analysis with only prebypass observations did not materially alter this relationship.
Using left ventricular dimensions measured by transesophageal echocardiography as the reference standard, this study also found that cardiac filling was better assessed by FTc than by PAD. Correlation coefficients between PAD and EDA were very poor. The correlation coefficients between FTc and EDA, although significantly better, were not especially strong by some criteria. In part, this finding may reflect the uncertainties caused by indirect methods of estimating chamber volume from time and pressure information.
The superiority of FTc as an index of left ventricular filling suggests that the EDM offers advantages over the PA catheter for cardiac monitoring. For PAD to increase, the left ventricle must be loaded sufficiently to raise left ventricular end-diastolic pressure. In a ventricle with normal diastolic compliance, major volume changes may not produce substantial pressure changes, thus limiting the usefulness of PAD as a marker for filling. Also, ventricular compliance can change after cardioplegic arrest, hypothermia, or ischemia, thereby making comparisons with prebypass values hazardous. In contrast, the FTc value implies the amount of time required to eject the stroke volume, corrected for heart rate. As the ventricle is filled, it requires more time to eject the greater stroke volume, even in ventricles with normal compliance. This may explain the superiority of FTc as an index of preload. Studies in volunteers subjected to maneuvers that alter preload, afterload, and contractility have documented good correlation of FTc with preload changes and minimal effect of alterations in inotropic state. Afterload reduction influences FTc and may need to be considered when assessing the volume status of a patient with esophageal Doppler techniques [19].
There are limitations to this study. To our knowledge, it is the first to compare EDM and PA catheter techniques, with a transit-time ultrasonic flow probe and transesophageal echocardiograms providing reference values. However, this additional instrumentation mandated that all measurements be made with an open chest and open pericardium. Sternal closure and positive-pressure ventilation may influence these results. The use of PA wedge pressure might prove to be a better reflection of left ventricular volume changes than PAD. We did not measure PA capillary wedge pressure because of concern about PA injury by cold catheters in the early period after bypass.
The study size is small (34 patients), and this could limit the ability to detect differences in some variables, particularly CO. In addition, patients with severely impaired ventricles were not included in this study, in part because of our reluctance to distend such hearts during the prebypass loading phase. Therefore, it may not be possible to generalize our results to this subpopulation. Patients included did not have severe peripheral vascular disease, aortic atherosclerosis, or valvular heart defects. The influence of these factors is not defined in this study. Finally, all data were collected by experienced personnel. Whether these results, particularly the EDM findings, can be replicated by observers less familiar with the technology remains to be established. Other investigators [20] have found EDM information to be reproducible after a brief learning curve.
Although many centers limit the use of any form of invasive monitoring to reduce costs or complications after cardiac operations, there is a group of patients in whom hemodynamic monitoring is important. On the basis of this study, the EDM may offer a less invasive technique for evaluating CO and a more accurate estimate of preload compared with the PA catheter. Further investigation of this technology appears warranted.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
W. Isakow and D. P. Schuster Extravascular lung water measurements and hemodynamic monitoring in the critically ill: bedside alternatives to the pulmonary artery catheter Am J Physiol Lung Cell Mol Physiol, December 1, 2006; 291(6): L1118 - L1131. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Collins, F. Girard, D. Boudreault, P. Chouinard, L. Normandin, P. Couture, M.-J. Caron, and M. Ruel Esophageal Doppler and thermodilution are not interchangeable for determination of cardiac output: [Le Doppler oesophagien et la thermodilution ne sont pas interchangeables pour preciser le debit cardiaque] Can J Anesth, November 1, 2005; 52(9): 978 - 985. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L. Raisis, K. J. Blissitt, W. Henley, K. Rogers, V. Adams, and L. E. Young The effects of halothane and isoflurane on cardiovascular function in laterally recumbent horses Br. J. Anaesth., September 1, 2005; 95(3): 317 - 325. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Iglesias, D. Bainbridge, and J. Murkin Intraoperative Echocardiography: Support for Decision Making in Cardiac Surgery Seminars in Cardiothoracic and Vascular Anesthesia, March 1, 2004; 8(1): 25 - 35. [Abstract] [PDF] |
||||
![]() |
T. Shiga, Z.'i. Wajima, T. Inoue, and R. Ogawa Nitrous oxide produces minimal hemodynamic changes in patients receiving a propofol-based anesthetic: an esophageal Doppler ultrasound study: [Le protoxyde d'azote produit des changements hemodynamiques minimes chez des patients qui recoivent une anesthesie a base de propofol : une etude de l'effet Doppler oesophagien] Can J Anesth, August 1, 2003; 50(7): 649 - 652. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Jacka, M. M. Cohen, T. To, J. H. Devitt, and R. Byrick The Use of and Preferences for the Transesophageal Echocardiogram and Pulmonary Artery Catheter Among Cardiovascular Anesthesiologists Anesth. Analg., May 1, 2002; 94(5): 1065 - 1071. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. B. Laupland and C. J. Bands Utility of esophageal Doppler as a minimally invasive hemodynamic monitor: a review: [Revue de l'utilite du Doppler oesophagien comme moniteur hemodynamique a effraction minimale] Can J Anesth, April 1, 2002; 49(4): 393 - 401. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |