|
|
||||||||
Ann Thorac Surg 1997;63:613-619
© 1997 The Society of Thoracic Surgeons
Department of Cardiac Surgery and Institute for Medical Statistics, Biometry, and Epidemiology, University Hospital Großhadern, Ludwig-Maximilians-University, Munich, Germany
Accepted for publication July 12, 1996.
| Abstract |
|---|
|
|
|---|
Methods. A retrospective follow-up of 87.1% of the patients representing 4,254 patient-years was completed. Surviving patients were examined by means of echocardiography.
Results. Survival rates after 10, 20, and 30 years were 62.3%, 39.4%, and 19.9% after aortic valve replacement and 75%, 36.5%, and 22.6% after mitral valve replacement (operative mortality excluded). Freedom from all valve-related complications, reoperations, and valve-related death was 66.4%, 43.3%, and 23.8% after aortic valve replacement and 73.4%, 35.4%, and 14.3% after mitral valve replacement. Of the surviving patients, 82% and 76% who received aortic or mitral valves, respectively, are in New York Heart Association class I or II. The pressure gradients of the aortic valves were between 20 and 73 mm Hg; those of the mitral valves were between 9 and 30 mm Hg. Fifty-two percent of aortic and 68% of mitral valves show no echocardiographic peculiarities. The left ventricular function in both groups is normal in 64%.
Conclusions. The long-term results together with the echocardiographic results show that after 30 years the Starr-Edwards valve represents a standard that still needs to be achieved by newer prostheses.
| Introduction |
|---|
|
|
|---|
Between April 1963 and May 1977, we used the Starr-Edwards prosthesis in our clinic for aortic and mitral valve replacements on a routine basis. This study reports 30 years of long-term results in these patients.
| Material and Methods |
|---|
|
|
|---|
|
Within the implantation time frame at our clinic, the design of the valve was changed several times by the manufacturer. The most noteworthy modifications were the covering of the cage with Dacron or Teflon cloth and the variation of the ball material. Table 1
describes the various models, manufacturing dates, the most important manufacturing changes, and the number of each model implanted at our clinic. In Figure 2
the most frequently used cloth-covered and noncloth-covered models in this study are shown.
|
|
Follow-up
All patients whose patient record did not note a date of death were sent a questionnaire to their last known address to be completed by the patients and their physician. A resting echocardiographic examination was performed on these identified survivors. Prosthetic function, peak transvalvular gradients, contractility, myocardial thickness, and the diameter of the atria and ventricles were determined.
Attempts were made to find patients who did not return a questionnaire by contacting central telephone information, resident registration offices, medical insurance companies, and finally known treating physicians. In patients who had already died, our patient records, the records of other treating physicians, and the records of hospitals and insurance companies were examined for the causes of death and nonfatal complications.
Those patients who could not be located are reported as being lost to follow-up. Retrospective postoperative follow-up for up to 30 years was possible for 241 patients (84.2%) in group A and 117 patients in group M (90.0%). Fifty-eight patients were lost to follow-up, mainly through relocation to foreign countries (Fig 3
). Overall follow-up was completed on 358 of 416 patients (86.1%), representing 4,254 patient years (pt-y) (group A, 2,649 pt-y; group M, 1,605 pt-y).
|
2 tests were performed to compare and to check the differences between patient groups A and M for significance. Hazard curves were computed to prove linearity of event rates. All statistical analyses were done using the computer program SPSS for Windows, Version 6.0, 1993 (SPSS Inc, Chicago, IL).
Definitions
All valve-related data have been collected and are reported according to the guidelines for reporting morbidity and mortality after cardiac valvular operations as published by Edmunds and associates [12].
| Results |
|---|
|
|
|---|
|
|
2 test (p > 0.10).
Complications
Thromboembolic events after the implantation of a Starr-Edwards aortic valve were observed 36 times, corresponding to a linearized rate of 1.36% per pt-y. Patients receiving a Starr-Edwards mitral valve suffered a thromboembolic complication in 20 cases, with a linearized rate of 1.25% per pt-y. Anticoagulant-related hemorrhage was recorded in 28 cases in group A (1.06% per pt-y) and in 9 cases in group M (0.56% per pt-y). According to the definition [12], structural valve deterioration has been found only once in the aortic group. In this prosthesis (model 1000) we found evidence of "ball variance" [13]. Nonstructural dysfunctions in aortic and mitral valves were seen 11 and 5 times, respectively. Endocarditis was diagnosed in 7 group A patients and 3 group M patients.
Reoperations were performed in a total of 33 patients (1.25% per pt-y or 13.7% of the patient population) with aortic prostheses and in 8 mitral valve patients (6.8% of the patient population). In 1 group A patient and 2 group M patients paravalvular leaks were closed. A new Starr-Edwards valve was implanted in 4 aortic patients and 1 mitral valve patient. Nine aortic valve patients and 1 mitral valve patient had a Starr-Edwards valve replaced with a bioprosthesis. Another type of mechanical valve was implanted in 17 group A and 4 group M patients. All patients whose original valve has been replaced have been censored from follow-up at the time of reoperation. The average length of time to reoperation was 8.2 years in group A and 9.7 years in group M. The reasons and number of reoperations are summarized in Table 3
. In four explanted group A valves and two group M valves, damage to the cloth covering of the cage was found. Only 2 of these patients, however, were reoperated on due to recurring emboli. A significant difference in the rates of reoperation between the various valve types was not found in either the aortic or mitral valve group.
|
|
|
|
Complications in Surviving Patients
Twenty surviving patients in group A and 4 surviving patients in group M required reoperation. The average time to reoperation in group A was 10.3 years. A new Starr-Edwards valve was implanted in 3 patients, and 17 patients received another type of artificial valve. For the mitral valve group, time to reoperation averaged 13.1 years. One patient again had a Starr-Edwards prosthesis implanted and 3 patients had other prostheses. Linearized complication rates based on pt-y in surviving patients (A, 1,192 pt-y; M, 749 pt-y) were calculated to provide a comparison with the entire patient population. Complication rates in all surviving patients are presented together with the total complications in Table 4
.
Surviving Patients With Starr-Edwards Prostheses
Follow-up examinations found 39 (16.4%) and 25 patients (21.4%) with their respective originally implanted Starr-Edwards aortic and mitral valves alive. Mean survival of these patients was 21.1 years in the aortic and 24.1 years in the mitral valve group. Six patients out of both groups have survived for 30 years. All patients are now anticoagulated, maintained with a prothrombin time between 20% and 30% (international normalized ratio, 4.5 to 3.0).
Thirty-two of the 39 patients (82%) in group A are currently in New York Heart Association class I or II, 7 patients are in class III, and no patients are in class IV. In group M, 19 of 25 patients (76%) are in class I or II, 6 patients (24%) in class III, and no patients are in class IV (Fig 7
). At the time of examination, all patients in class III were under close medical supervision. Twenty-four patients with aortic prostheses (61.5%) and 16 patients (64%) with mitral prostheses are retired; 15 and 9, respectively, are working.
|
Echocardiographic Results
The peak pressure gradients ranged between 20 and 73 mm Hg (mean, 36.2 mm Hg) in aortic prostheses and between 9 and 30 mm Hg (mean, 16.6 mm Hg) in the mitral prostheses. Grade 1 insufficiency was identified in 33% of aortic and 20% of mitral valves. A stenotic condition with high gradients was seen in 15% of group A and 12% of group M patients.
Left ventricular myocardial function was measured as a function of fractional shortening and was normal in 64% of both groups. Significant left ventricular hypertrophy was seen in 64% of group A patients (wall thickness > 12 mm). In group M, left ventricular configuration was normal in 52% of the patients. The left atrium was normal in 62% of patients with an aortic prosthesis; however, left atrial dilatation was found in all patients with mitral prostheses (left atrial diameter > 40 mm). All echocardiographic results are summarized in Table 5
.
|
| Comment |
|---|
|
|
|---|
The survival rates of our patients are higher than those of other authors [4, 6, 9, 10, 14, 15] within the reported time periods after Starr-Edwards valve replacement. In these studies the 10-year survival rates after aortic valve replacement range between 54% [10] and 56% [9] and between 54% [10] and 70% [6] after mitral valve replacement. After 15 years, the survival of patients who received an aortic valve was between 40% [6] and 42% [10]; that of mitral valve recipients was between 35% [10] and 48% [14]. The higher survival rates in our patient groups are caused by excluding early mortality. The high 30-day mortality that we observed early on is probably related to the limited possibilities of myocardial protection and the still-developing possibilities of postoperative intensive care medicine and not to the prosthesis itself. Therefore it is justifiable not to include early mortality in the evaluation of the 30-year long-term qualities of this device. When operative mortality is still included, the survival in our patient group is 49.2% (A) and 70.0% (M) after 10 years, 46.6% (A) and 39.2% after 20 years, and 16.6 (A) and 18.4% (M) at 30 years. Now survival in our aortic group up to 15 years is slightly less than in the compared studies; survival in the mitral group does not differ from the other authors' reports.
Considering that this is a retrospective study, there may have been incomplete data collection, particularly for complications. This is especially true because we did this review only once at the end of 30 years and not on an annual or semiannual basis. To rule out this possible data loss as much as possible, we additionally evaluated the complications of surviving patients, where we were certain to obtain all information relevant to specific questions. If one places these data next to the complication rates of the total patient population (see Table 4
), there only is a clear difference in the rates of thromboembolism in the aortic group and in the incidence of reoperations in both groups. As is expected, among the survivors there are more patients who have undergone reoperation and who have probably survived as a result of the reoperation. The lower incidence of thromboembolism implies that this could be the most frequent cause of valve-related death in our patients, and therefore the rate of thromboembolism in surviving patients will be less than in the total group. One major reason for this low complication rate may be the mean patient age of 40.1 years, which is much younger than in most series. Because patient characteristics are known to have a profound influence on valve performance, this additionally must be considered in comparing our data with other studies with an older patient population.
The better long-term survival rate of mitral valve patients when compared with aortic valve patients partially contradicts other experiences, but is in our opinion related to operative techniques. In our patients the myocardium was not perfused during aortic valve replacement, whereas in mitral valve procedures continual myocardial perfusion with blood was maintained. This should result in a reduction of intraoperative myocardial injury, which is positively revealed in the long-term results.
The fact that the majority of the surviving patients are in New York Heart Association class II and that more than one-third of these patients are still working demonstrates a good quality of life in patients who have had a Starr-Edwards valve for more than 20 years. No patient complained about the noise of the prosthesis; most of them state that they do not even hear it.
Echocardiographic results show that over time the heart does adapt in specific conditions to the prosthesis. Fifty-four percent of hypertrophied left ventricles with only 15% stenotic valves in the surviving aortic patient group prove that sufficient cardiac function could only be maintained by significant ventricular hypertrophy. Therefore, not every hypertrophy is necessarily be related to a nonstructural valvular dysfunction like entrapment by pannus or suture [10]; hypertrophy may be caused by the high transvalvular gradient of this particular type of prosthesis itself. Dilatation of the left atrium in the same way is the response to the additional burden of a mitral Starr-Edwards prosthesis. Because in patients with mitral valve replacement, the aortic valves are not affected, the left ventricles are not hypertrophied or dilated and demonstrate normal function. Overall, the echocardiographic results are satisfactory, and we believe that earlier implanted Starr-Edwards valves should not be prophylatically removed, eg, after 20 years. The removal of the prosthesis should be evaluated on an individual basis in patients who require an additional cardiac operation for other indications.
In conclusion, in view of the ultra-long-term course, echocardiographic findings, and the quality of life of the surviving patients, we conclude that the results with Starr-Edwards prostheses are satisfactory. Because only two of the valve models used in this study are still available today, these results do not demonstrate a final standard for succeeding devices; nevertheless, until the modern prostheses can show a significant improvement in long-term results, the Starr-Edwards valve will represent the gold standard in mechanical valve replacement.
| Footnotes |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. Gudbjartsson, T. Absi, and S. Aranki Mitral Valve Replacement Card. Surg. Adult, January 1, 2008; 3(2008): 1031 - 1068. [Full Text] |
||||
![]() |
J. L. Svennevig, M. Abdelnoor, and S. Nitter-Hauge Twenty-Five Year Experience With the Medtronic-Hall Valve Prosthesis in the Aortic Position: A Follow-Up Cohort Study of 816 Consecutive Patients Circulation, October 16, 2007; 116(16): 1795 - 1800. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. S. De Santo, G. Romano, A. D. Corte, F. Tizzano, A. Petraio, C. Amarelli, M. De Feo, G. Dialetto, M. Scardone, and M. Cotrufo Mitral mechanical replacement in young rheumatic women: Analysis of long-term survival, valve-related complications, and pregnancy outcomes over a 3707-patient-year follow-up J. Thorac. Cardiovasc. Surg., July 1, 2005; 130(1): 13 - 19. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Shiono, Y. Sezai, A. Sezai, M. Hata, M. Iida, and N. Negishi Long-Term Results of the Cloth-Covered Starr-Edwards Ball Valve Ann. Thorac. Surg., July 1, 2005; 80(1): 204 - 209. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Wu, R. Gregorio, A. Renzulli, F. Onorati, M. De Feo, G. Grunkemeier, and M. Cotrufo Mechanical heart valves: Are two leaflets better than one? J. Thorac. Cardiovasc. Surg., April 1, 2004; 127(4): 1171 - 1179. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Higashita, S. Ichikawa, H. Niinami, T. Ban, Y. Suda, and Y. Takeuchi Long-term results after Starr-Edwards mitral valve replacement in children aged 5 years or younger Ann. Thorac. Surg., March 1, 2003; 75(3): 826 - 829. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Gudbjartsson, S. Aranki, and L. H. Cohn Mechanical/Bioprosthetic Mitral Valve Replacement Card. Surg. Adult, January 1, 2003; 2(2003): 951 - 986. [Full Text] |
||||
![]() |
C. Detter, T. Fischlein, C. Feldmeier, G. Nollert, and B. Reichart Aortic valvotomy for congenital valvular aortic stenosis: a 37-year experience Ann. Thorac. Surg., May 1, 2001; 71(5): 1564 - 1571. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Lund, H. K. Pilegaard, L. B. Ilkjaer, S. L. Nielsen, and H. Arildsen Performance profile of the Starr-Edwards aortic cloth covered valve, track valve, and silastic ball valve Eur. J. Cardiothorac. Surg., October 1, 1999; 16(4): 403 - 413. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |