Ann Thorac Surg 2001;71:S147-S149
© 2001 The Society of Thoracic Surgeons
a Division of Cardiothoracic Surgery, Department of Surgery, University of Louisville, Louisville, Kentucky, USA
b Jewish Hospital Heart and Lung Institute, Louisville, Kentucky, USA
Address reprint requests to Dr Dowling, Division of Cardiothoracic Surgery, University of Louisville, 201 Abraham Flexner Way, Suite 1200, Louisville, KY 40202
e-mail: rddowl01{at}athena.louisville.edu
Presented at the Fifth International Conference on Circulatory Support Devices for Severe Cardiac Failure, New York, NY, Sept 1517, 2000.
Abstract
Background. The AbioCor implantable replacement heart (IRH) has been developed as an alternative to transplant (ie, destination therapy). We report our experience with the AbioCor IRH in a bovine model at the University of Louisville.
Methods. Male Holstein cows were used (85 to 115 kg). The internal controller, battery, and secondary transcutaneous energy transfer coil were implanted in the right flank. After cardiopulmonary bypass, the thoracic unit was implanted orthotopically. After removal of air and weaning from cardiopulmonary bypass, the AbioCor was connected to internal components and energy transfer through transcutaneous energy transfer coils was achieved.
Results. Nineteen animals underwent implantation of the AbioCor IRH for a proposed 30-day duration. There were 6 deaths, none related to device malfunction. All animals demonstrated normal hemodynamics with normal pressures in the aorta, pulmonary artery, left atrium, and right atrium. There was no significant hemolysis and all animals demonstrated normal end organ function. The internal battery allowed for brief periods of untethered mobility.
Conclusions. The AbioCor IRH has resulted in normal hemodynamics and normal end organ function without evidence of hemolysis in a bovine model.
This article has been cited by other articles:
![]() |
R. D. Dowling, A. S. Ghaly, and L. A. Gray Jr Creation of a diaphragm patch to facilitate placement of the AbioCor implantable replacement heart Ann. Thorac. Surg., May 1, 2004; 77(5): 1849 - 1850. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Dowling, L. A. Gray Jr, S. W. Etoch, H. Laks, D. Marelli, L. Samuels, J. Entwistle, G. Couper, G. J. Vlahakes, and O. H. Frazier Initial experience with the AbioCor Implantable Replacement Heart System J. Thorac. Cardiovasc. Surg., January 1, 2004; 127(1): 131 - 141. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. Myers, K. Robertson, T. Pool, N. Shah, I. Gregoric, and O. H. Frazier Continuous flow pumps and total artificial hearts: management issues Ann. Thorac. Surg., June 1, 2003; 75(90060): S79 - 85. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. L. Kukuy, M. C. Oz, and Y. Naka Long-Term Mechanical Circulatory Support Card. Surg. Adult, January 1, 2003; 2(2003): 1491 - 1506. [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |