Ann Thorac Surg 1995;59:689-694
© 1995 The Society of Thoracic Surgeons
Divisions of Cardiovascular and Thoracic Surgery and Pediatric Cardiology, University of Minnesota, Minneapolis, Minnesota
Accepted for publication November 23, 1994.
A series of 14 infants and small children ranging from 7 months to 7 years in age (mean, 2.5 years) underwent implantation of transvenous pacemaker systems. Three factors are of utmost importance in children: small subclavian vein size, thin subcutaneous layer in the chest, and growth. A five-point protocol is followed strictly: (1) duplex assessment of upper veins, (2) use of active fixation leads, (3) use of short (36 to 45 cm) leads, (4) anchoring of pulse generator with nonabsorbable material to prevent migration, and (5) routine use of the ``lateral approach'' in children more than 2 years old when the pulse generator is implanted in the chest. Because lead diameters measure 2 to 2.3 mm, a one-lead system needs a vein diameter of 5 mm (cross-sectional area of 19 mm2). A two-lead system needs a vein at least 7 mm in diameter and a cross-sectional area of 38 mm2 to prevent vein occlusion. Therefore all children less than 3 years of age had the leads implanted via the internal jugular vein. In 50% of children between 4 and 7 years of age, the internal jugular system also was used. Children more than 7 years old have leads implanted via the subclavian veins. Duplex ultrasound assessment of the upper veins is important to decide route of implantation. Use of short leads is recommended to reduce bulk at the pulse generator site. The ``lateral approach'' prevents problems at the generator implantation site.
This article has been cited by other articles:
![]() |
M. S. Silvetti, F. Drago, S. Marcora, and L. Rava Outcome of single-chamber, ventricular pacemakers with transvenous leads implanted in children Europace, October 1, 2007; 9(10): 894 - 899. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Silvetti, F. Drago, A. De Santis, G. Grutter, L. Rava, L. Monti, and R. Fruhwirth Single-centre experience on endocardial and epicardial pacemaker system function in neonates and infants Europace, June 1, 2007; 9(6): 426 - 431. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Agarwal, G. S. Krishnan, S. Abraham, K. Bhatt, P. Sekar, S. Kulkarni, and K. M. Cherian Extrapleural Intrathoracic Implantation of Permanent Pacemaker in the Pediatric Age Group Ann. Thorac. Surg., April 1, 2007; 83(4): 1549 - 1552. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Schreiber, M. Kostolny, A. Eicken, and R. Lange Nonthoracotomy Cardioverter Defibrillator Implantation in a 2-Year-Old Infant With Long QT Syndrome Ann. Thorac. Surg., June 1, 2006; 81(6): e27 - e28. [Abstract] [Full Text] [PDF] |
||||
![]() |
F U. Ten Cate, J Breur, N Boramanand, J Crosson, A Friedman, J Brenner, E Meijboom, and N Sreeram Endocardial and epicardial steroid lead pacing in the neonatal and paediatric age group Heart, October 1, 2002; 88(4): 392 - 396. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. G. Warner, D. P. Halpin, C. I. Berul, and D. D. Payne Placement of a permanent epicardial pacemaker in children using a subcostal approach Ann. Thorac. Surg., July 1, 1999; 68(1): 173 - 175. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Molina Undertreatment and Overtreatment of Patients With Infected Antiarrhythmic Implantable Devices Ann. Thorac. Surg., February 1, 1997; 63(2): 504 - 509. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |