The Annals of Thoracic Surgery, Vol 52, 815-820, Copyright © 1991 by The Society of Thoracic Surgeons
M Butterfield, J Fisher, GA Davies and TJ Spyt
The hydrodynamic function of each size of the CarboMedics valve was
assessed in a pulsatile flow simulator. The mean pressure difference with
respect to forward flow, regurgitant volumes, and total energy loss across
each valve were analyzed. The results for the 23-mm aortic and 29-mm mitral
CarboMedics valves were compared with those for the St. Jude Medical and
Bjork-Shiley Monostrut valves. Results showed good hydrodynamic function
for each CarboMedics valve, although the pressure difference and total
energy loss across the 19-mm aortic valve was high. The hydrodynamic
function of the CarboMedics valve was comparable with that of the St. Jude
Medical valve. Both valves showed similar leakage volumes, which were
significantly larger than that for the Bjork-Shiley Monostrut valve. On
account of this the total energy loss in the Bjork-Shiley valve was
significantly less than that for the bileaflet valves in the aortic
position. Concern remains for the continuing presence of high closed-valve
regurgitation in the bileaflet valves.
ARTICLES
Comparative study of the hydrodynamic function of the CarboMedics valve
Cardiac Research Unit, Killingbeck Hospital, Leeds, England.
This article has been cited by other articles:
![]() |
T. Gudbjartsson, T. Absi, and S. Aranki Mitral Valve Replacement Card. Surg. Adult, January 1, 2008; 3(2008): 1031 - 1068. [Full Text] |
||||
![]() |
C. H. Kang, H. Ahn, K. H. Kim, and K.-B. Kim Long-Term Result of 1144 CarboMedics Mechanical Valve Implantations Ann. Thorac. Surg., June 1, 2005; 79(6): 1939 - 1944. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hartrumpf, J. M. Albes, T. Krempl, V. Rudolph, and T. Wahlers The hemodynamic performance of standard bileaflet valves is impaired by a tilted implantation position Eur. J. Cardiothorac. Surg., March 1, 2003; 23(3): 283 - 291. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Gudbjartsson, S. Aranki, and L. H. Cohn Mechanical/Bioprosthetic Mitral Valve Replacement Card. Surg. Adult, January 1, 2003; 2(2003): 951 - 986. [Full Text] |
||||
![]() |
C. Bartels, R. G. Leyh, J.F. M. Bechtel, E. Joubert-Hubner, and H.-H. Sievers Discrepancies Between Sizer and Valve Dimensions: Implications for Small Aortic Root Ann. Thorac. Surg., June 1, 1998; 65(6): 1631 - 1633. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Bernal, J. M. Rabasa, F. Gutierrez-Garcia, C. Morales, J. F. Nistal, and J. M. Revuelta The CarboMedics Valve: Experience With 1,049 Implants Ann. Thorac. Surg., January 1, 1998; 65(1): 137 - 143. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini COMPARISON OF HEMODYNAMIC PERFORMANCES OF ST. JUDE MEDICAL AND CARBOMEDICS 21 MM AORTIC PROSTHESES BY MEANS OF DOBUTAMINE STRESS ECHOCARDIOGRAPHY J. Thorac. Cardiovasc. Surg., February 1, 1996; 111(2): 408 - 415. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |