The Annals of Thoracic Surgery, Vol 45, 210-212, Copyright © 1988 by The Society of Thoracic Surgeons
RG Masters and WJ Keon
Three cases of acute extrinsic obstruction of the Medtronic-Hall prosthesis
in the mitral position are described. In 2 cases papillary muscle and in 1
case suture overhang caused sticking of the tilting disk against the valve
rim. The mechanisms of extrinsic obstruction of the tilting disk valve, its
recognition, and its prevention are discussed with a review of the
literature.
ARTICLES
Extrinsic obstruction of the Medtronic-Hall disk valve in the mitral position
Division of Cardiothoracic Surgery, University of Ottawa Heart Institute, Ontario, Canada.
This article has been cited by other articles:
![]() |
M. Ruel, R. B. Streeter, R. de la Torre, J. R. Liddicoat, and W. E. Cohn Intracardiac ultrasonic suture welding for knotless mitral valve replacement Eur. J. Cardiothorac. Surg., February 1, 2002; 21(2): 245 - 248. [Abstract] [Full Text] [PDF] |
||||