GP Pai, RG Ellison, JW Rubin, HV Moore and MV Kamath
During the last fourteen years, 377 unileaflet tilting-disc prosthetic
valves (Bjork-Shiley and Medtronic Hall) have been used for single or
multiple valve replacements with and without concomitant coronary artery
by-pass grafting. In the past five years, five instances of disc
immobilization (three in the mitral and two in the aortic position)
occurred either at the time of weaning from cardiopulmonary bypass or
immediately thereafter. When the implanted site of the prosthetic valve was
the mitral position, reexploration in 2 patients revealed chordal remnants
in the subannular area stuck between the disc occluder and the valve ring,
thereby immobilizing the disc. In the third instance, the free movement of
the disc was impeded by the left ventricular myocardium. In the aortic
position, an unraveled suture impacted between the disc occluder and the
valve ring immobilized the disc in 1 patient. In the other patient, the
cause of the malfunction could not be determined at the time of
reexploration. The 1 death among these 5 patients was directly related to
the malfunction of the prosthesis. The mechanism, recognition, treatment,
and prevention of this catastrophic malfunction of tilting-disc valves are
discussed.
ARTICLES
Disc immobilization of Bjork-Shiley and Medtronic Hall valves during and immediately after valve replacement
This article has been cited by other articles:
![]() |
L. S. Rallidis, C. Papadopoulos, J. Kanakakis, I. Paraskevaidis, and D. T. Kremastinos 'Obstruction alternans' of a Sorin tilting disc prosthetic mitral valve: a case of emergency action Eur J Echocardiogr, September 1, 2008; 9(5): 704 - 706. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Machler, G. Reiter, M. Perthel, U. Reiter, P. Bergmann, M. Zink, R. Rienmuller, and J. Laas Influence of a tilting prosthetic mitral valve orientation on the left ventricular flow -- an experimental in vivo magnetic resonance imaging study Eur. J. Cardiothorac. Surg., July 1, 2007; 32(1): 102 - 107. [Abstract] [Full Text] [PDF] |
||||
![]() |
W Keeble and S M Cobbe Pressure damping, a "billowing" septum, and an eerie silence: perioperative, intermittent obstruction of a mitral valve prosthesis Heart, September 1, 2000; 84 (3): e6 - e6. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |