The Annals of Thoracic Surgery, Vol 41, 155-157, Copyright © 1986 by The Society of Thoracic Surgeons
GA Lamas, EM Antman, JP Gold, NS Braunwald and JJ Collins
Both electrocautery and internal defibrillation are routinely used in
cardiac operations. The cases of 5 patients are presented in whom
backup-mode reversion or injury of permanently implanted pulse generators
occurred during a cardiac procedure. The theoretical explanations for
backup-mode reversion and generator or tissue injury are discussed, and
recommendations are made for the management of patients with a pacemaker
who are to undergo a cardiac operation.
ARTICLES
Pacemaker backup-mode reversion and injury during cardiac surgery
This article has been cited by other articles:
![]() |
M. Ozeren, O.V. Dogan, C. Duzgun, and E. Yucel Use of an ultrasonic scalpel in the open-heart reoperation of a patient with pacemaker Eur. J. Cardiothorac. Surg., April 1, 2002; 21(4): 761 - 762. [Abstract] [Full Text] [PDF] |
||||