Of 836 patients with a mitral or aortic bioprosthesis who were followed
over an 8-year period, 32 required reoperation for prosthetic dysfunction.
Of the 26 prostheses removed at operation, 18 (12 mitral and 6 aortic
prostheses) showed tears in the valve leaflets. In 7 of the 12 patients
with rupture of "mitral" leaflets and in 1 of the 6 patients with torn
"aortic" leaflets, the onset of symptoms was sudden, requiring urgent
reoperation. Two of the 18 patients died early after reoperation; both had
progressive disability and underwent elective operation. In 5 of the 7
patients with torn leaflets in a mitral prosthesis who had acute onset of
symptoms, a single tear was found and no gross evidence of calcification
was detected. Macroscopic calcification was found in only 8 of the 18
(44.4%) prostheses with torn leaflets (mitral and aortic). Microscopic
examination revealed calcific deposits in 9 of the 12 mitral and in all of
the aortic bioprostheses. Of the 54 leaflets examined, 29 had tears: 10
were type I; 6, type II; 7, type III; and 6, type IV. Tears in porcine
valve leaflets are not an uncommon cause of dysfunction. These tears often
lead to acute hemodynamic changes, particularly in the mitral position, and
usually require emergency reoperation. They may occur in prostheses that
appear grossly normal.
ARTICLES
Late tears in leaflets of porcine bioprostheses in adults
This article has been cited by other articles:
![]() |
P. Bertolini, G. B. Luciani, B. Vecchi, P. Pugliese, and A. Mazzucco Aortic valve replacement with the biocor PSB stentless xenograft Ann. Thorac. Surg., August 1, 1998; 66(2): 425 - 430. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. B. Luciani, P. Bertolini, B. Vecchi, and A. Mazzucco Midterm results after aortic valve replacement with freehand stentless xenografts: a comparison of three prostheses J. Thorac. Cardiovasc. Surg., June 1, 1998; 115(6): 1287 - 1291. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Dunmore-Buyze, D. R. Boughner, N. Macris, and I. Vesely A COMPARISON OF MACROSCOPIC LIPID CONTENT WITHIN PORCINE PULMONARY AND AORTIC VALVES: Implications for bioprosthetic valves J. Thorac. Cardiovasc. Surg., December 1, 1995; 110(6): 1756 - 1761. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |