The Annals of Thoracic Surgery, Vol 23, 374-375, Copyright © 1977 by The Society of Thoracic Surgeons
DF Larson, B Mayall and RM Anderson
A chemical contaminant found on the interior surface of the extracorporeal
apparatus manufactured by the William Harvey Corporation was believed to
have caused severe hematuria in 2 children. The biological activity of the
contaminant was tested by mixing 50 ml of whole blood in contaminated
cardiotomies; after 5 minutes of exposure lysis of one-sixth of the red
blood cells was found. This hemolytic agent was concluded to be bisphenol A
dissolved in an organic solvent, but the mechanism of contamination was
never found.
ARTICLES
Hemolysis due to chemical contamination of clinical perfusion apparatus
This article has been cited by other articles:
![]() |
A. D. Papaconstantinou, T. H. Umbreit, B. R. Fisher, P. L. Goering, N. T. Lappas, and K. M. Brown Bisphenol A-Induced Increase in Uterine Weight and Alterations in Uterine Morphology in Ovariectomized B6C3F1 Mice: Role of the Estrogen Receptor Toxicol. Sci., August 1, 2000; 56(2): 332 - 339. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |